Matrices listed below were spiked with certain level of MAPtau and the recovery rates were calculated by comparing the measured value to the expected amount of MAPtau in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-102
93
EDTA plasma(n=5)
85-100
93
heparin plasma(n=5)
88-93
90
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MAPtau and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
MAPtau ELISA Kit allows for the in vitro quantitative determination of MAPtau concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%