Matrices listed below were spiked with certain level of COL4 and the recovery rates were calculated by comparing the measured value to the expected amount of COL4 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
89-104
98
EDTA plasma(n=5)
85-98
92
heparin plasma(n=5)
89-101
97
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of COL4 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
COL4 ELISA Kit allows for the in vitro quantitative determination of COL4 concentrations in serum, plasma, tissue homogenates and other biological fluids.