CAM ELISA Kit, Calmodulin ELISA Kit, Calcium-modulated protein ELISA Kit
Reactivity
Rat
Detection Method
Double Antibody
Detection Range
0.156-10ng/ml
Sample Type
Cell lysate, Tissue lysate, Other liquid samples
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of CAM and the recovery rates were calculated by comparing the measured value to the expected amount of CAM in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-102
94
EDTA plasma(n=5)
91-105
99
heparin plasma(n=5)
87-100
92
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CAM and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
CAM ELISA Kit allows for the in vitro quantitative determination of CAM concentrations in serum, plasma, tissue homogenates and other biological fluids.