Matrices listed below were spiked with certain level of ARNT2 and the recovery rates were calculated by comparing the measured value to the expected amount of ARNT2 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
94-105
99
EDTA plasma(n=5)
92-101
98
heparin plasma(n=5)
87-101
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ARNT2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
ARNT2/Aryl hydrocarbon receptor nuclear translocator 2/ARNT protein 2
Notes
For Research Use Only
UniProt ID
Q78E60
Antigen
ARNT2
Applications
ELISA
Applications Notes
ARNT2 ELISA Kit allows for the in vitro quantitative determination of ARNT2 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months
Reviews of Rat ARNT2(Aryl Hydrocarbon Receptor Nuclear Translocator 2) ELISA Kit