Apo D ELISA Kit, APOD ELISA Kit, apolipoprotein D ELISA Kit
Reactivity
Rabbit
Detection Method
Double Antibody
Detection Range
3.906-250ng/ml
Sample Type
Other liquid samples, Serum, Cell culture supernatants, Plasma, Cell lysate
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of ApoD and the recovery rates were calculated by comparing the measured value to the expected amount of ApoD in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
95-105
100
EDTA plasma(n=5)
89-103
95
heparin plasma(n=5)
86-104
96
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ApoD and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
ApoD ELISA Kit allows for the in vitro quantitative determination of ApoD concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months