Other liquid samples, Serum, Cell culture supernatants, Plasma, Cell lysate
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of IL-19 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-19 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
87-101
96
EDTA plasma(n=5)
88-105
98
heparin plasma(n=5)
85-96
90
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-19 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
86-104%
95-100%
87-100%
EDTA plasma(n=5)
86-98%
85-93%
83-96%
heparin plasma(n=5)
80-98%
81-100%
81-96%
Sensitivity
< 18.75pg/ml
Notes
For Research Use Only
UniProt ID
Q8CJ70
Applications
ELISA
Applications Notes
IL-19 ELISA Kit allows for the in vitro quantitative determination of IL-19 concentrations in serum, plasma, tissue homogenates and other biological fluids.