Other liquid samples, Serum, Cell culture supernatants, Plasma, Cell lysate
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of IL-13 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-13 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
86-104
94
EDTA plasma(n=5)
86-101
94
heparin plasma(n=5)
85-101
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-13 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Interleukin-13,T-cell activation protein P600,Il-13
Notes
For Research Use Only
UniProt ID
P20109
Antigen
IL-13
Applications
ELISA
Applications Notes
IL-13 ELISA Kit allows for the in vitro quantitative determination of IL-13 concentrations in serum, plasma, tissue homogenates and other biological fluids.