Matrices listed below were spiked with certain level of Comp and the recovery rates were calculated by comparing the measured value to the expected amount of Comp in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-104
94
EDTA plasma(n=5)
94-101
97
heparin plasma(n=5)
87-104
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Comp and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Comp ELISA Kit allows for the in vitro quantitative determination of Comp concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months
Reviews of Mouse Comp(Cartilage oligomeric matrix protein) ELISA Kit