Matrices listed below were spiked with certain level of anti-SARS-CoV2 and the recovery rates were calculated by comparing the measured value to the expected amount of anti-SARS-CoV2 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
92-101
96
EDTA plasma(n=5)
89-105
97
heparin plasma(n=5)
91-101
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of anti-SARS-CoV2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
88-102%
87-103%
87-102%
EDTA plasma(n=5)
88-95%
86-96%
82-95%
heparin plasma(n=5)
81-98%
82-90%
84-99%
Sensitivity
< 0.938ng/ml
Notes
For Research Use Only
Applications
ELISA
Applications Notes
anti-SARS-CoV2IgG ELISA Kit allows for the in vitro quantitative determination of anti-SARS-CoV2 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months