Matrices listed below were spiked with certain level of CD30 and the recovery rates were calculated by comparing the measured value to the expected amount of CD30 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
93-103
98
EDTA plasma(n=5)
86-104
95
heparin plasma(n=5)
86-105
98
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CD30 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
85-95%
92-104%
91-103%
EDTA plasma(n=5)
89-101%
87-98%
85-99%
heparin plasma(n=5)
85-99%
87-99%
85-100%
Sensitivity
< 18.75pg/ml
Notes
For Research Use Only
UniProt ID
A0A2K5VW07
Applications
ELISA
Applications Notes
CD30 ELISA Kit allows for the in vitro quantitative determination of CD30 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%