Matrices listed below were spiked with certain level of ZNF804A and the recovery rates were calculated by comparing the measured value to the expected amount of ZNF804A in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-103
93
EDTA plasma(n=5)
87-104
95
heparin plasma(n=5)
93-102
98
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ZNF804A and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
89-102%
91-102%
86-99%
EDTA plasma(n=5)
84-101%
82-89%
89-100%
heparin plasma(n=5)
84-98%
86-95%
84-99%
Sensitivity
< 9.375pg/ml
Notes
For Research Use Only
UniProt ID
Q7Z570
Applications
ELISA
Applications Notes
ZNF804A ELISA Kit allows for the in vitro quantitative determination of ZNF804A concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months