Matrices listed below were spiked with certain level of SOD and the recovery rates were calculated by comparing the measured value to the expected amount of SOD in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
92-104
96
EDTA plasma(n=5)
87-103
99
heparin plasma(n=5)
89-97
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SOD and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
85-102%
85-100%
94-105%
EDTA plasma(n=5)
83-93%
87-99%
86-100%
heparin plasma(n=5)
85-98%
80-96%
92-99%
Sensitivity
< 0.469ng/ml
Notes
For Research Use Only
Antigen
SOD
Applications
ELISA
Applications Notes
total SOD ELISA Kit allows for the in vitro quantitative determination of SOD concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months