Other liquid samples, Serum, Cell culture supernatants, Plasma, Cell lysate
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of sdLDL and the recovery rates were calculated by comparing the measured value to the expected amount of sdLDL in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
86-105
96
EDTA plasma(n=5)
87-105
94
heparin plasma(n=5)
85-103
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sdLDL and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
sdLDL ELISA Kit allows for the in vitro quantitative determination of sdLDL concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months