Matrices listed below were spiked with certain level of PLSCR2 and the recovery rates were calculated by comparing the measured value to the expected amount of PLSCR2 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
91-105
98
EDTA plasma(n=5)
89-102
95
heparin plasma(n=5)
85-102
95
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PLSCR2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
PLSCR2 ELISA Kit allows for the in vitro quantitative determination of PLSCR2 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months