Matrices listed below were spiked with certain level of PLA2G4D and the recovery rates were calculated by comparing the measured value to the expected amount of PLA2G4D in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
86-102
95
EDTA plasma(n=5)
90-100
97
heparin plasma(n=5)
85-100
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PLA2G4D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
PLA2G4D ELISA Kit allows for the in vitro quantitative determination of PLA2G4D concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months