Matrices listed below were spiked with certain level of Phospho Tau and the recovery rates were calculated by comparing the measured value to the expected amount of Phospho Tau in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
89-105
97
EDTA plasma(n=5)
90-100
96
heparin plasma(n=5)
90-103
95
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Phospho Tau and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
89-101%
89-99%
91-102%
EDTA plasma(n=5)
91-100%
89-99%
92-100%
heparin plasma(n=5)
95-96%
90-99%
90-97%
Sensitivity
< 4.688pg/ml
Notes
For Research Use Only
UniProt ID
P10636
Antigen
Phospho Tau(P181)
Applications
ELISA
Applications Notes
Phospho Tau ELISA Kit allows for the in vitro quantitative determination of Phospho Tau concentrations in serum, plasma, tissue homogenates and other biological fluids.