Matrices listed below were spiked with certain level of OVA sIgE and the recovery rates were calculated by comparing the measured value to the expected amount of OVA sIgE in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-105
93
EDTA plasma(n=5)
92-101
96
heparin plasma(n=5)
85-101
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of OVA sIgE and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
OVA sIgE ELISA Kit allows for the in vitro quantitative determination of OVA sIgE concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months