Matrices listed below were spiked with certain level of LOXL4 and the recovery rates were calculated by comparing the measured value to the expected amount of LOXL4 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
86-105
96
EDTA plasma(n=5)
86-104
95
heparin plasma(n=5)
85-102
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LOXL4 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
LOXL4/Lysyl oxidase-like protein 4/Lysyl oxidase homolog 4
Notes
For Research Use Only
UniProt ID
Q96JB6
Antigen
LOXL4
Applications
ELISA
Applications Notes
LOXL4 ELISA Kit allows for the in vitro quantitative determination of LOXL4 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months