Matrices listed below were spiked with certain level of LDL and the recovery rates were calculated by comparing the measured value to the expected amount of LDL in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-105
98
EDTA plasma(n=5)
86-104
92
heparin plasma(n=5)
86-101
94
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LDL and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
LDL ELISA Kit allows for the in vitro quantitative determination of LDL concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months