Matrices listed below were spiked with certain level of GLUD1 and the recovery rates were calculated by comparing the measured value to the expected amount of GLUD1 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
86-103
98
EDTA plasma(n=5)
90-103
98
heparin plasma(n=5)
86-102
91
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GLUD1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
GLUD1 ELISA Kit allows for the in vitro quantitative determination of GLUD1 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months
Reviews of Human GLUD1(Glutamate dehydrogenase 1, mitochondrial) ELISA Kit