Matrices listed below were spiked with certain level of DSP and the recovery rates were calculated by comparing the measured value to the expected amount of DSP in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-98
95
EDTA plasma(n=5)
90-100
95
heparin plasma(n=5)
92-103
100
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of DSP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
DSP ELISA Kit allows for the in vitro quantitative determination of DSP concentrations in serum, plasma, tissue homogenates and other biological fluids.