Matrices listed below were spiked with certain level of DCP and the recovery rates were calculated by comparing the measured value to the expected amount of DCP in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
85-105
94
EDTA plasma(n=5)
86-101
94
heparin plasma(n=5)
91-96
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of DCP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
DCP ELISA Kit allows for the in vitro quantitative determination of DCP concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months