Matrices listed below were spiked with certain level of Bcl2L12 and the recovery rates were calculated by comparing the measured value to the expected amount of Bcl2L12 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
87-103
94
EDTA plasma(n=5)
85-99
92
heparin plasma(n=5)
88-102
93
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Bcl2L12 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Bcl2L2 ELISA Kit allows for the in vitro quantitative determination of Bcl2L12 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months