Matrices listed below were spiked with certain level of anti-IL18BP and the recovery rates were calculated by comparing the measured value to the expected amount of anti-IL18BP in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
87-105
100
EDTA plasma(n=5)
94-101
97
heparin plasma(n=5)
85-96
91
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of anti-IL18BP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
86-104%
87-105%
86-103%
EDTA plasma(n=5)
90-98%
89-97%
86-95%
heparin plasma(n=5)
86-99%
86-100%
88-99%
Sensitivity
< 1.875ng/ml
Notes
For Research Use Only
UniProt ID
O95998
Applications
ELISA
Applications Notes
anti-IL18BPantibody Elisa kit allows for the in vitro quantitative determination of anti-IL18BP concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage
2-8 °C for 6 months
Reviews of Human anti-IL18BP(Interleukin 18 Binding Protein)antibody Elisa kit