Other liquid samples, Serum, Cell culture supernatants, Plasma, Cell lysate
Assay Type
Sandwich
Recovery
Matrices listed below were spiked with certain level of IL-2 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-2 in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-100
93
EDTA plasma(n=5)
85-104
94
heparin plasma(n=5)
89-99
92
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
IL-2 ELISA Kit allows for the in vitro quantitative determination of IL-2 concentrations in serum, plasma, tissue homogenates and other biological fluids.