Matrices listed below were spiked with certain level of H3cit-NE and the recovery rates were calculated by comparing the measured value to the expected amount of H3cit-NE in samples.
Matrix
Recovery range(%)
Average(%)
serum(n=5)
88-102
94
EDTA plasma(n=5)
92-102
98
heparin plasma(n=5)
88-105
96
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of H3cit-NE and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample
1:2
1:4
1:8
serum(n=5)
91-105%
87-100%
89-104%
EDTA plasma(n=5)
87-98%
88-101%
85-101%
heparin plasma(n=5)
92-100%
87-99%
85-98%
Sensitivity
< 0.094ng/ml
Notes
For Research Use Only
Applications
ELISA
Applications Notes
H3cit-NE ELISA kit allows for the in vitro quantitative determination of H3cit-NE concentrations in serum, plasma, tissue homogenates and other biological fluids.
Precision
Intra-Assay: CV<8% Inter-Assay: CV<10%
Storage